Mandatory Calibration and Tests for SmokeStack Testing Equipment (Malaysian Standard MS 1596 and MS 1723)
Doc ID: AERHQWW-ecy3-en
Last revised: 0, 23.07.06
This page contains a summary of the mandatory calibration (and tests/checks/verifications) required for equipment used in Malaysian Standard MS 1596 and MS 1723 for smokestack testing ("stack testing" and isokinetic sampling on stationary sources of air pollution).
For smokestack testing (and especially isokinetic sampling), the sampling procedures are more important than the analytical procedures. This is because of the difficulty in getting a representative sample of the flue gas inside the smokestack. The pollutants could be in solid, liquid, or gaseous states, or a mixture (e.g., solid in gas) which is not well mixed, and with many other interferences.
We strongly recommend attaching proof of the below calibrations and tests/checks to the final smokestack testing report.
Notes:
- Please note that this page will be focusing on the more uncommon calibrations and tests, which are special to smokestack testing.
- This page is based on what is written in MS 1596:2003 and MS 1723:2003. Refer to local authorities for actual local rules!
- MS 1596:2003 and MS 1723:2003 are only for general particulate matter "dust", and not any specific chemical (in any state: solid, liquid, or gaseous)
- MS 1596:2003 is based on ISO 9096:1992, and ISO 10780. Refer to the "Foreword" of the Standard.
- As stated in section 10.1 of MS 1723:2003, a [smoke-stack] test shall be done a minimum of three times, to ensure reasonable accuracies and make it possible to compare results.
- This is the same for US EPA Methods, where 1 stack "test" = 3 sampling "runs"
- Refer also to the Environmental Quality Act (Clean Air) Regulations 2014, Regulation 23
Other useful links:
- Aer Sampling Basic Training Video with names of equipment and their individual components
- Click here to order Malaysian Standards at the Department of Standards
- To order ISO Standards, click here to visit ISO online store
- Why is calibration important? Click here to link to YouTube videos:
- To return to Aer Sampling's Smokestack Testing Encyclopedia content page, click here, or scan the QR Code below:
- Images of entire sampling train of equipment for MS 1596:2003 (Method i, dry gas measurement)
- Table of mandatory calibrations
- Integrating Dry Gas Meter (DGM)
- Critical Orifices (to calibrate DGM)
- Temperature Sensor
- Thermocouple Simulator (to calibrate temperature sensor)
- Field Balance
- Analytical Balance
- Orifice Meters, Rotameters
- Manometers and Pressure Gages
- Calipers (to test S-type Pitot Tubes)
- Angle Indicator or Inclinometers (to test S-type Pitot Tubes)
- Profile Projector or equivalent (to test S-type Pitot Tubes)
- Flue Gas Analyzers
- Timer
- Internal Diameter Micrometers (to test Sampling Nozzle)
- Other Analytical Instruments
- Rod to measure internal diameter of smoke-stack
- *Table of mandatory tests/verifications
- Leak Check - Pitot Line
- Leak Check - Sampling Gas Line
- Leak Check - Meter Console (Vacuum side, "Front-Half")
- Leak Check - Meter Console (Pressure side, "Back-Half")
- S-type Pitot Tube - in a Combination Probe
- S-type Pitot Tube - by itself
- Suction Nozzle - Entry Wall Thickness
- Suction Nozzle - Diameter "Size"
- Optional Calibrations
- Barometer
- S-type Pitot Tubes
- L-type (Standard) Pitot tubes
Table of mandatory calibrations (click individual item for details):
- Including calibrations to ensure metrological traceability
What to calibrate |
Frequency |
Calibrated Against |
1. Integrating Dry Gas Meter (DGM), inside the Meter Console
|
Once a year (minimum)
Refer to:
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors
Recommend to use Critical Orifices |
2. Critical Orifice (to calibrate DGM)
|
Once a year (minimum)
Refer to:
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors
Recommend to use Bell Gas Provers
Click here (or scan QR Code below) to view Aer Sampling's traceable calibration service for Critical Orifices |
3. Temperature Sensor, inside the Meter Console
Need to calibrate every thermocouple connected to Temperature Sensor too (via all the thermocouple plugs and sockets) |
Once a year (minimum)
Refer to:
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors
Recommend to use Thermocouple Simulators
|
4. Thermocouple Simulator
Used to calibrate the Temperature Sensor
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors
Recommended: Minimum once a year
Refer to:
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors |
5. Field Balance
Used for weighing contents of impingers
|
Once a year (minimum)
Refer to:
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors
|
6. Analytical Balance
Used to measure weight of filter papers (MS 1596:2003: "particle separator")
|
Once a year (minimum)
Refer to:
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors |
7. Flow rate measuring devices
|
Once a year (minimum)
Refer to:
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors
Recommended to use Critical Orifices, refer to section below
|
8. Manometers and Pressure Gages
Used to measure differential pressure head "Delta P" of flue gas inside smokestack, which is then used to calculate velocity |
Once a year (minimum)
Refer to:
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors
|
9. Calipers
This is used to test S-type Pitot Tubes construction specifications (lengths and widths), to avoid having to calibrate the S-type Pitot Tubes.
Refer to:
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors
Recommended: Minimum once a year
Refer to:
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors |
10. Angle Indicator or Inclinometer, with Bull's Eye Level
These are used to test S-type Pitot Tubes construction specifications (angles), to avoid having to calibrate.
Refer to:
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors
Recommended: Minimum once a year
Refer to:
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors |
11. Profile Projector "Optical Comparator", or Microscope with image analysis software, or equivalent
This is used to test S-type Pitot Tubes construction specifications; dimension L1 and L2 of MS 1596:2003, Figure D1.
Refer to:
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors
Recommended: Minimum once a year
Refer to:
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors |
12. Flue Gas Analyzers
|
Once a year (minimum)
Refer to:
|
Not mentioned (but usually calibration gases of specified and traceable purity)
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors
Recommended standards:
|
13. Timer
Used to countdown to sampling at the next traverse points inside the smokestack |
Once a year (minimum)
Refer to:
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors |
14. Internal Diameter Micrometers
Used to test entry diameter of Sampling Nozzles, which are used to determine flow rate etc.
Refer to:
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors
Recommended: Minimum once a year
Refer to:
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors |
15. Other Analytical Instruments Not mentioned in MS 1596:2003 and MS 1723:2003, because these Standards are for "dust" only
Some examples:
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors
Recommended: Minimum once a year
Refer to:
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors |
16. Rod to measure internal diameter of smokestack
Refer to:
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors
Recommended: Minimum once a year
Refer to:
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors |
Related topics: ISO/IEC 17025, ILAC MRA, scope of accreditation, stationary/fixed sources of air pollution, SAMM
------
List of mandatory tests/verifications (click individual item for details):
What to test/verify
|
Frequency
|
Test Using
|
1. Leak Check: Pitot Line
The "Pitot Line" consist of three items: Pitot Tubes, Pitot Tube connections (Umbilical Cable), Manometers or Pressure Gages (inside Meter Console)
Refer to:
|
Before sampling starts (after equipment assembled)
Refer to:
--
After each period of use (each stack test)
Refer to:
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors
Recommend to use Manometers or Pressure Gages, both in meter console |
2. Leak Check: Sampled Gas Line
From the Suction Nozzle to the pump valves and fittings, and Dry Gas Meter (DGM) inside the Meter Console
Leak rate shall be less than 1% of the volumetric flow rate at a vacuum of at least 50 kPa (14.765 inHg)
Refer to:
|
Before start of sampling (after assembling entire sampling train, from the nozzle to meter console)
Refer to:
Recommend to check for leaks after the sampling too |
Dry Gas Meter (DGM), usually inside the Meter Console |
3. Leak Check - Meter Console (Vacuum Side, "Front-Half")
From the Sampled Gas inlet to the pump valves and fittings inside the Meter Console
|
When necessary (This is part of the sampling gas line train leak check, but just for the meter console) |
Dry Gas Meter (DGM), which is inside the Meter Console
Recommend to use procedures in US EPA Method 5
|
4. Leak Check - Meter Console (Pressure Side, "Back-Half")
From the pump valves and fittings to the Orifice Meter (downstream after the DGM)
|
When necessary (This is part of the sampling gas line train leak check, but just for the meter console) |
Manometers or Pressure Gages); Analog or Digital), which are inside the Meter Console
Recommend to use procedures in US EPA Method 5
|
5. Suction Nozzle - Entry wall thickness
If the wall thickness at the mouth of the Nozzle is too thick, the diameter of the nozzle is calculated using a specified formula.
Refer to:
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors
Recommended: Before each use, to test if nozzle still in original shape, given the importance of the nozzle in determining isokinetic velocity
Refer to:
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors
Recommended: Vernier Calipers |
6. Suction Nozzle - Diameter "Size"
The diameter of the sampling nozzle where the flue gas enters the nozzle. This directly impacts the isokinetic rate calculations.
Refer to:
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors
Recommended: Before each use, to test if nozzle size is still valid, and not dented or damaged in any way, because you would have to select the correct nozzle size to sample isokinetically
Refer to:
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors
Recommended: Internal Diameter Micrometers |
7. S-Type Pitot Tube - as part of a Combination Probe
Combination Probes are probes where the suction nozzle, S-type pitot tubes, and temperature sensor are combined for sampling.
Construction dimensions to test and assign S-type Pitot Tube Calibration Factor ("Coefficient", "Calibration Constant") of 0.84. If failed any tests below, replace or repair pitot tube, or calibrate using Wind Tunnel (together with the Combination Probe), per MS 1596:2003 Appendix D. For clearer procedures, we recommend using procedures in US EPA Method 2, Section 10.1 and 10.1.1.
MS 1596:2003 Appendix D (Refer to ISO 10780 for more details):
|
After use at each field site
Refer to:
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors
Recommended to use:
|
8. S-Type Pitot Tube - Without Probe
Pitot Tubes by itself, without nozzles, thermocouple etc.
Construction dimensions to test and assign S-type Pitot Tube Calibration Factor ("Coefficient", "Calibration Constant") of 0.84. If failed any tests below, replace or repair pitot tube, or calibrate using Wind Tunnel, per MS 1596:2003 Appendix D. For clearer procedures, we recommend using procedures in US EPA Method 2, Section 10.1 and 10.1.1.
MS 1596:2003 Appendix D (Refer to ISO 10780 for more details)
|
After use at each field site
Refer to:
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors
Recommended to use:
|
Optional Calibrations
These are calibrations that are not mentioned inside Malaysian Standards MS 1596:2003 and MS 1723:2003, and are not used to calibrate other instruments.
What to Calibrate
|
Frequency
|
Calibrated Against
|
1. Barometer
Used to measure barometric pressure, which is used in isokinetic rate calculations |
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors
Recommended: Minimum once a year
Refer to:
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors
Recommended: Mercury Barometer
|
2. Pitot Tubes (S-type)
If failed tests on constructions specifications, a calibration in a Wind Tunnel against a Standard (L-type) Pitot Tube (MS 1596:2003: "Standard Pitot Static Tube"), is required.
Refer to:
|
Prior to use
Refer to:
---
After use at each field site (if it fails construction specification tests)
Refer to:
---
Once a year (minimum)
Refer to:
|
Standard (L-type) Pitot Tube in a Wind Tunnel
Refer to:
|
3. Pitot Tubes (Standard; L-type)
Other names:
|
Once a year (minimum)
Refer to:
|
Not mentioned
Refer to Jabatan Alam Sekitar or ISO/IEC 17025 Technical Assessors |
-----
Information presented here is subjected to Aer Sampling Group's terms and conditions.
In case of discrepancy between English version and the non-English version, the English version shall prevail.